17
Technical Information
Use
Input/Output Menu
This menu contains parameters that determine the operating mode of
the inputs and outputs of the controller.
Figure 41
270/306
WATER COOLING: indicates whether the pump is cooled with
water or not. The controller limits the power supplied to the
pump according to this setting.
ACTIVE STOP: enables (YES) or disables (NO) active braking.
STOP SPEED READING: enables (YES) or disables (NO) the
pump rotational frequency reading function also during the
slowing phase that follows a stop command
IMPUT / OPTPUT
Flow 3/6
Turbo-V 81-AG Rack Controller User Manual / 87-900-986-01
START - INTERLOCK
START MODE
START/STOP LEVEL
START/STOP EDGE
INTERLOCK TYPE
CONTINUOUS
IMPULSIVE
ANALOG OUT TYPE
FREQUENCY
POWER
CURRENT
PRESSURE
RUNNING
START OUT MODE
STARTING
SPEED DISPLAY
KRPM
Hz
LCD CONTRAST
0
7
Flow 3/6
(I)