Turbo-V 81-AG Rack Controller User Manual / 87-900-986-01
Range
Description
B0= Stop Speed Reading
This function measures the pump speed after a
stop command.
0 = Off
1 = On
B1= Active Stop
This function brakes the pump after a stop
command, by using the motor as a generator.
0 = Off
1 = On
B2 - Reserved
B3-4 = Start/Stop/Interlock Mode
Change the logic of Start/Stop and Interlock
inputs.
Ask to Agilent personnel for this option
B0 = Vent & Purge Auto/Man
0 = Auto (the controller drives purge and vent
valves automatically)
1 = Manual (the user can command the valves by
serial line or profibus)
B1 = Gas Load Type
Select here the gas load
0 = Ar
1 = N2
B2 – Interlock Type
0 = continuos (the interlock must be always
present to run the pump)
1 = at start (the interlock must be present only at
start command)
B3 = Setpoint R1 Logic
0 = active close (NO)
1 = active open (NC)