J7 – Vent
Figure 7
This is dedicated 24 Vdc connector to control the optional vent valve
(see orderable parts table). The vent valve can be driven by controller
automatically or by serial line. If you already have a vent valve
installed 110 Vac powered, this valve must be replaced by a new
model (see orderable parts table).
Vent Valve
CLOSED
Figure 8
Turbo-V 81-AG Rack Controller User Manual / 87-900-986-01
Vent Connector
OPEN
STOP
RAMP
START COMMAND
Vent Valve diagram in "Auto" mode
Technical Information
Controller Connection
Vent Valve Open
Vent Valve Relay
NORMAL
Win. 126
STOP COMMAND
(I)
Time Win. 147
215/306
17