Input Buffer
Tab. 20
Offset
Size
Unit Res
Byte
Byte
0
1
V
1
1
°C
2
1
°C
3
1
°C
4
1
-
Turbo-V 81-AG Rack Controller User Manual / 87-900-986-01
Range
Description
0-100
Voltage
Voltage provided to the motor
0-100
Pump Bearing Temperature
Read the pump upper bearing temperature
If > 60 °C the controller goes into fail status and
the pump is stopped
0-100
Controller Power Supply Temperature
Read the controller power section temperature if >
75 °C the controller goes in fail status "controller
over-temperature"
0-100
Controller CPU Section Temperature
Read the controller CPU section temperature if >
80 °C the controller goes in fail "controller over-
temperature"
0-6
B0-3 – Status
0 = Stop
1= Waiting for interlock (interlock connections on
remote I/O connector missing)
2 = Ramp (the pump is increasing the speed)
3 = Auto tuning (the speed has been reduced
because the gas load is higher than the Power
Limit Applied)
4 = Braking (the pump is reducing the speed using
the motor)
5 = Normal operation (the pump is at target
speed)
6 = Fail (see error code byte Input Buffer Byte-27)