Parameter Buffer
Tab. 18
Offset
Size
Unit Res
Byte
Byte
0
1
-
1
2
mA/Hz/sec
3
2
Sec
5
2
mA/Hz/sec
7
2
Sec
9
2
0.2 sec/bit
Turbo-V 81-AG Rack Controller User Manual / 87-900-986-01
Range
Description
0-99
Gauge gas load type
The value is divided by 10 (entering 12 the
controller will apply 1.2)
Enter here the correction factor for gas (necessary
only if gauge is connected).
0-10000
Setpoint 1 value
Enter here the threshold value for the R1 output,
valid only if B0-B2 of Parameter Buffer Offset-20 is
equal to 0,1,2 or 3.
0-30000
Setpoint 1 Mask Time
Set here the time, measured from the start pump
event, while the R1 output is masked (not
activated)
0-10000
Setpoint 2 value
Enter here the threshold value for the R2 output,
valid only if B3-B5 of Parameter Buffer Offset-20 is
equal to 0,1,2 or 3.
0-30000
Setpoint 2 Mask Time
Set here the time, measured from the start pump
event, while the R2 output is masked (not
activated)
0-30000
Vent Valve Opening Delay
Set the delay time between the stop command
and vent opening. Valid only in "purge/vent auto
mode" (Parameter Buffer, Offset 19, B0 = 0). 0 =
infinite (the valve will never open).
Technical Information
Profibus Option