6. Create a subcutaneous tunnel between the midline incision and the exit site until the straw is visible
and accessible at the exit point.
7. Unscrew and remove the tunneling tool handle.
8. Grasp the tip of the tool with one hand while holding the straw in place with the other hand. Pull the
tunneling tool shaft out through the straw.
9. Push the lead or extension proximal ends through the straw, then withdraw the straw.
10. For Permanent Trials, proceed to "Connecting the Lead Extension" on page 10.
11. For Permanent IPG Implantation, if extensions are used, proceed to the instructions for "Connecting
the Lead Extension" on page 10.
12. For Permanent IPG Implantation, refer to instructions for connecting to the IPG in the appropriate DFU
for your SCS system as listed in your Reference Guide.
Note: The following Codman Disposable Catheter Passers may be used in place of the Boston Scientific
tunneling tool for the 2x8 Surgical Lead only:
REF 82-1515 (36 cm); REF 82-1516 (55 cm); REF 82-1517 (65 cm)
Note: When using a Codman Disposable Catheter Passer, tunnel from the IPG pocket to the midline
incision to the using the standard technique.
Instructions for Use
Surgical Leads Directions for Use
90970892-17 Rev A 9 of 277