WARNING!
To ensure proper lubrication of the bearings the pump has a default setting
called soft start. If the pump has been sitting for an extended period or on
its initial startup the controller will automatically start in soft start. This
mode of operation brings the product up to speed over an extended amount
of time.
This automatic operating mode cannot be deactivated by the user. The user
can possibly decide to start the product always in "soft start mode", acting
on an appropriate serial command, as specified below.
The "Soft Start" mode allows the correct redistribution of the grease present in
the bearings, after a long period of inactivity.
The "Soft Start" mode is automatically activated in the following cases:
1
2
WARNING!
This mode can be activated manually using the serial command WIN100.
The WIN 100 has no effect on the automatic soft start at the first starting
and after 60 days from the last moment in FULL SPEED.
See "Windows Protocol" section.
Cleaning
The exterior surface of the 305-IC may be cleaned with mild detergent only.
222/302
At the first starting;
After 60 days from reaching the last moment in "FULL SPEED" condition.
TwisTorr 305-IC User Manual / 87-901-058-01