Window Description
WIN # READ
WRITE
0
Always
Serial
1
Always
Serial
8
Always
Always
100
Always
Stop & Serial
101
Always
Always
102
Always
Always
103
Always
Always
104
Always
Always
105
Always
Always
108
Always
Always
109
Always
Always
110
Always
Always
111
Always
Always
117
Always
Always
120
Always
Always
121
Always
Always
122
Always
Always
125
Always
Always
126
Always
Always
TwisTorr 305-IC User Manual / 87-901-058-01
TYPE
NAME
L
Start/Stop
L
Low Speed
A
Serial/Remote
L
Soft Start
N
Setpoint Type
N
Setpoint Value
N
Setpoint Delay
L
Setpoint Logic
N
Setpoint histeresis
N
Baud Rate
L
Pump Life Reset
L
Interlock Type
N
Analog Output Type
N
Low Speed Adjust Frequency
N
Target Frequency
N
Maximum frequency
L
Set vent valve
A
Set the vent valve operation
N
Vent valve opening delay
Technical Information
Window Description
MEANING
Start/Stop command
Low Speed command
Configures controller
Enable/Disable Soft Start option
It defines set point type:
0 = Freqency
1 = Power
2 = Time
3 = Normal
It defines the threshold value for Set Point
It defines the setpoint mask time from start
It defines the functionally logic. "High level active"
indicates an output ON when the signal is above the
threshold and "Low level active" is the opposite.
It defines the percentage value of Set Point Hysteresis
It defines the serial communication baud rate:
0 = 600 bit/s
1 = 1200 bit/s
2 = 2400 bit/s
3 = 4800 bit/s
4 = 9600 bit/s
5 = 19200 bit/s
6 = 38400 bit/s
Pump Life Reset
It defines the interlock type. The interlock can be checked
at start or continuously
It defines the analog output type:
0 = Frequency (10V = OMEGA_MAX_ABS)
1 = Power (10V = P_RANGE)
2 = Body Temperature (10V = 125°C)
It defines the Low Speed frequency value
It defines the target excitation frequency
It defines the maximum frequency settable by WIN120
Set vent valve
Set the vent valve operation
0 = Automatic
1 = On command
Vent valve opening delay (expressed in 0.2sec)
8