Error Rectification - Ivoclar Vivadent Bluephase PowerCure Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 38
6. What if ...?
Indicator
Causes
The device is overheated.
Red "x"
lights up
Electronic component of the
handpiece is defective.
Battery empty
Red "x" and
battery symbol
light up
Battery contacts dirty
In the 3sCure program, the curing time
3s flashes
is set to 3 seconds and cannot be
changed. The 3sCure program can be
used no more than twice in direct
succession.
– Power pack not connected or
The charging base
defective
is not illuminated
during charging
– Battery fully charged
18

Error rectification

Allow the device to cool down and try again after a
certain time. If the error persists, please contact your
dealer or your local Service Centre.
Remove and reinsert the battery. If the error persits,
please contact your dealer or your local Service
Centre.
Place the device in the charging base and charge it.
Remove battery and clean the battery contacts.
After two consecutive curing cycles, an extended
waiting time of 30 seconds is necessary for safety
reasons before the next curing cycle can be started.
If for compelling reasons an additional curing cycle is
required before the prescribed interval of 30 seconds
has elapsed, the 3sCure program can be activated
again by selecting it anew with the program selection
button.
Check if the power pack is correctly positioned in
the charging base or if the power pack is connected
to the power supply by means of the power cord.

Publicité

Table des Matières
loading

Table des Matières