Guidance and manufacturer's declaration – electromagnetic immunity
Bluephase PowerCure is intended for use in the electromagnetic environment specified below. The customer or the user of Bluephase PowerCure
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 V
rms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Table: According to IEC60601-1-2
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
Recommended separation distances between portable and mobile RF communications equipment
and "Bluephase Style PowerCure"
Bluephase PowerCure is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the
user of Bluephase PowerCure can help prevent electromagnetic interference by maintaining a miminum distance between portable and mobile RF
communications equipment (transmitters) and Bluephase PowerCure as recommended below, according to the maximum out-put power of the
communication equipment.
Separation distance according to frequency
Rated maximum
output power
of transmitter
150 kHz to
W
80 MHz
d = 0.35 √P
0.01
0.035
0.1
1
10
100
Table: According to IEC60601-1-2
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Compliance level
10 V
10 V/m
of transmitter
m
80 MHz to
800 MHz
d = 0.35 √P
0.12
0.11
0.38
0.35
1.2
1.1
3.8
3.5
12
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should not be used
closer to any part of Bluephase PowerCure, including cables, than the
recommended separation distance caculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 0.35 √P
d = 0.35 √P 80 MHz to 800 MHz
d = 0.70 √P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the trans mitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electro-
magnetic site survey,
b
frequency range.
Interference may occur in the vicinity of equipment marked with the
following symbol:
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which Bluephase PowerCure
is used exceeds the applicable RF compliance level above, Bluephase PowerCure should
be observed to verify normal operation. If abnormal performance is ob-served, additional
measures may be necessary, such as reorienting or relocating Bluephase PowerCure.
b
Over the frequency range 150 kHz to 80 MHz, field strength should be less than 10 V/m.
800 MHz to
2.5 GHz
d = 0.7 √P
0.23
0.73
2.3
7.3
23
a
should be less than the compliance level in each
111