Ivoclar Vivadent Bluephase PowerCure Mode D'emploi page 14

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 38
Observe the Instructions for Use of the material
applied when selecting the curing time and intensity.
The curing recommendations for composite materials
apply to all shades and, if not mentioned otherwise in
the Instructions for Use, to a maximum layer thickness
of 2 mm. Generally, these re commendations apply to
situations where the emission window of the light guide
is placed directly over the material to be polymerized.
Increasing the distance between the light source and the
material will require the curing time to be extended
Curing times
Tetric PowerFill / Tetric PowerFlow
4 mm
3sCure System
Adhese Universal
Composites
• 2 mm
IPS Empress Direct /
IPS Empress Direct Effect
Restorative
Tetric EvoCeram / Tetric EvoFlow
materials
• 4 mm
Tetric EvoCeram Bulk Fill /
Tetric EvoFlow Bulk Fill
Variolink Esthetic LC
Variolink Esthetic DC
Indirect restorations /
cementation
materials
Multilink Automix
Telio CS Link
Temporary materials
Telio CS Inlay/Onlay
Helioseal F Plus
Miscellaneous
IPS Empress Direct Color
IPS Empress Direct Opaque
MultiCore Flow / Multicore HB
1) The 3sCure program may only be used on the occlusal surfaces of Class I & II
restorations in the posterior region and may not be used at all in cases of caries
profunda, or deep cavities.
2) Applies to a maximum layer thickness of 2 mm and provided that the Instructions
for Use of the respective material do not state any other recommendation
(which may be the case, e.g. with certain dentin shades)
14
2)
3)
4)
/
5)
/ SpeedCEM Plus
5)
accordingly. For instance, if the distance to the material
is approx. 8 mm, the effective light output is reduced by
approx. 50%. In this case, the recommended curing time
has to be doubled.
1) The information provided herein applies to the
10>9 mm light guide supplied in the delivery form.
2) The information regarding heat development and
burn hazards must be taken into consideration
(see Safety notes).
HIGH POWER
1,200 mW/cm
2
10 seconds
10 seconds
10 seconds
10 seconds
per mm ceramic:
10 seconds per surface
per mm ceramic:
5)
20 seconds per surface
10 seconds per surface
10 seconds
10 seconds
20 seconds
20 seconds
20 seconds
3) Applies to a maximum layer thickness of 4 mm and provided that the Instructions
for Use of the respective material do not state any other recommendation
(which may be the case, e.g. with certain dentin shades)
4) Applies to light-curing
5) Applies to optional light-curing
TURBO
2,000 mW/cm
2
5 seconds
––
5 seconds
5 seconds
per mm ceramic:
5 seconds per surface
per mm ceramic:
2 x 5 seconds
per surface
10 seconds per surface
5 seconds
––
5 seconds
2 x 5 seconds
2 x 5 seconds
3sCure
3,000 mW/cm
2
3 seconds
1)
3 seconds
1)
––
––
––
––
––
––
––