DK50 PLUS
18.1.
Maintenance intervals
once
Time interval
a day
Product operational
checks
Drain condensate from the
air tank
)**
- At high humidity
Drain condensate from the
air tank
)**
- At normal humidity
Product function check
Leak check connections
and inspect device
Inspection of electrical
connections
Cooler and fan check
Filter element replacement
in the dryer filter
Filter element replacement
in theKJF-1
Safety valve check
Conduct a "repeated test"
per EN 62353
Aggregate inlet filter
replacement
)*
)* data is presented in hours; if not possible, then the data is considered in years
)** only for compressors without dryers
02/2018
once
once
once
2000
every 2
a week
a year
hours
years
x
x
x
x
x
x
x
x
x
x
x
x
- 27 -
4000
6000
8000
10000
12000
hours
hours
hours
hours
hours
x
x
Set of spare
Chapter
parts
18.2
-
18.5
-
18.5
-
8
-
18.3
-
18.4
-
18.10
-
18.8
025200304-000
18.9
025200061-000
18.6
-
18
-
x
18.7
025200126-000
NP-DK50 PLUS-1_01-2018-MD A
Performed by
user
user
user
user
qualified
professional
qualified
professional
qualified
professional
qualified
professional
qualified
professional
qualified
professional
qualified
professional
qualified
professional