Fault
... the display "Er:X:11"
appears?
... the display "Er:X:12"
appears?
... the display "Er:X:16"
appears?
... when the Enter key
(send/print) is pressed,
the "Er:x:71" display
appears?
... when the Enter key
(send/print) is pressed,
the "Er:x:72" display
appears?
11.SERVICING
76 •
Cause/Remedy
The scale has too high a load, or too great a load
is applied in one corner.
- Reduce load on scale or distribute weight
more evenly.
- Restart scale.
The scale has been switched on with too great a
load.
- Remove load.
- Restart scale.
Oscillations have been induced in the scale, pre-
venting determination of the zero point.
- Do not touch the trolley or table supporting
the scale.
- Restart scale.
Data transmission not possible, wireless module is
disabled.
- Activate wireless module ("Activating the wire-
less module (SYS)" on page 68).
Data transmission is not possible, no wireless
group set up.
- Set up wireless group ("Setting up a wireless
group (Lrn)" on page 69).
On leaving the factory, your seca scale has an accu-
racy of ±0,15 % or better. To preserve this level of
accuracy, the product must be set up with care and
serviced regularly. Depending on how frequently the
scale is used, we recommended servicing at intervals
of 3 to 5 years.
NOTICE!
Incorrect measurements as a result of poor
servicing
– Have servicing and repairs carried out
exclusively by seca service or by an
authorized service partner.
– You can find service partners in your area at
www.seca.com or by sending an e-mail to
service@seca.com.