Servicing/Recalibration; Information On Maintenance And Recalibration; Checking The Content Of The Calibration Counter - Seca 645 Mode D'emploi

Masquer les pouces Voir aussi pour 645:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 113
Problem
The display "Er:X:16"
appears.
The Enter key
(send/print) is pressed
and the display "Er:X:71"
appears.
The Enter key
(send/print) is pressed
and the display
"Er:X:72" appears.

10.SERVICING/RECALIBRATION

10.1 Information on maintenance and recalibration

10.2 Checking the content of the calibration counter

Cause / Solution
The scale was caused to oscillate, the zero point
could not be determined.
- Restart the scale.
No data transmission possible, wireless module
deactivated.
- Activate wireless module (see "Activating the
wireless module (SYS)" on page 99).
No data transmission possible, no wireless group
set up.
- Set up wireless group (see "Setting up a
wireless group (Lrn)" on page 99).
We recommend having your device serviced prior to
recalibration.
NOTICE!
Incorrect measurements as a result of poor
servicing
Have servicing and repairs carried out
exclusively by seca service or by an
authorized service partner.
You can find service partners in your area at
www.seca.com or by sending an e-mail to
service@seca.com.
Have an authorized technician perform recalibration
according to national legal regulations.
Recalibration is necessary whenever one or more calib-
ration seals are damaged or the contents of the calibra-
tion counter no longer match the number on the
applicable calibration counter sticker. If calibration
seals are damaged, contact seca service directly.
This seca scale is calibrated. Calibrations may only be
performed by authorized agencies. To guarantee this,
the scale is equipped with a calibration counter to
record any change in calibration-related data.
Servicing/recalibration • 107
English

Publicité

Table des Matières
loading

Table des Matières