8. TROUBLESHOOTING
Problem
No weight is displayed with a
load on the weighing platform
0.0 does not appear before
weighing
- - - - - appears before
weighing
The weight display flashes
One display segment is
displayed continuously or not
at all
is displayed
bAtt is displayed
StOP is displayed
Er:[No.]:11 is displayed
Er:[No.]:12 is displayed
Er:[No.]:16 is displayed
Er:[No.]:32 is displayed
9. SERVICING
28 •
Cause
The device has no power supply
A load was already on the device before it
was switched on
A load was already on the device before it
was switched on
Consistent weight value has not yet been
detected
The display is defective
Battery voltage is dropping
The batteries have discharged
Maximum capacity has been exceeded
The device has too high a load or too high
a load in one corner
The device has been switched on with too
high a load
The device was caused to oscillate, the
zero point could not be determined
An internal communication error has
occurred
The product must be set up carefully and serviced regularly. Depending on
how frequently the product is used, we recommended servicing at intervals of
3 to 5 years.
CAUTION!
Incorrect measurements as a result of poor servicing
► Have servicing and repairs carried out exclusively by seca Service or
by an authorized service partner.
► You can find service partners in your area at www.seca.com or by
sending an e-mail to service@seca.com.
Remedy
• Switch on the measuring station
• In the case of battery operation: Insert
batteries
• In the case of power supply operation:
Connect the plug-in power supply unit
• Take the load off the measuring station
• Switch off the measuring station, then
switch it back on again
• Take the load off the measuring station
• Switch off the measuring station, then
switch it back on again
• One of the following functions is active:
,
,
. Wait until the mea-
Hold
Autohold
Tare
suring station has detected a consistent
weight value. The display stops flashing.
• Take the load off the measuring station,
wait until
is displayed and weigh again
0.0
• Inform seca Service
• Change the batteries
• Change the batteries
• Take the load off the measuring station
• Remove the load from the measuring sta-
tion or distribute the weight more evenly
• Switch off the measuring station, then
switch it back on again
• If the problem persists, inform
seca Service
• Take the load off the measuring station
• Switch off the measuring station, then
switch it back on again
• Switch off the measuring station, then
switch it back on again
• Switch off the measuring station, then
switch it back on again
• If the problem persists, inform
seca Service