Problem
The display "Er:[No.]:16"
appears.
The Enter key (send/
print) is pressed and the
display "Er:[No.]:71"
appears.
The Enter key (send/
print) is pressed and the
display "Er:[No.]:72"
appears.
10.SERVICING/VERIFICATION
10.1 Information on servicing and verification
10.2 Checking the content of the verification counter
Cause / Solution
The scale was caused to oscillate, the zero point
could not be determined.
- Restart the scale
No data transmission possible, wireless module
deactivated.
- Activate wireless module (see "Activating the
wireless module (SYS)" on page 95)
No data transmission possible, no wireless group
set up.
- Set up wireless group (see "Setting up a wire-
less group (Lrn)" on page 95)
We recommend having your device serviced prior to
verification.
NOTICE!
Incorrect measurements as a result of poor
servicing
►
Have servicing and repairs carried out
exclusively by seca Service or by an
authorized service partner.
►
You can find service partners in your area at
www.seca.com or by sending an e-mail to
service@seca.com.
Have an authorized technician perform verification
according to national legal regulations.
Verification is necessary whenever one or more verifica-
tion seals are damaged or the contents of the verifica-
tion counter no longer match the number on the
applicable verification counter sticker. If verification
seals are damaged, contact seca Service directly.
This seca scale is verified. Verifications may only be
performed by authorized agencies. To guarantee this,
the scale is equipped with a verification counter to
record any change in verification-related data.
Servicing/Verification • 103