SMRT CHARGER TROUBLESHOOTING (CONTINUED)
English
Problem/Failure Recommended Action
The SMRT
Charger
indicator
LEDs are NOT
illuminated
when the Pak
is inserted
Return To Table of Contents
1-30
Troubleshooting
1.
Unplug the charger, wait five seconds, and plug the
charger back into the wall outlet.
2.
Verify that the charger power LED is illuminated.
A. If not, see the "SMRT Charger power LED is not
illuminated" section.
3.
Verify that the Pak is fully inserted into the charger.
A. If not, reinsert the Pak fully, listen for an audible
click and verify the charger indicator LEDs.
4.
Try another Pak to identify any Pak issues.
Note: Separate and label Paks during troubleshooting.
A. If the charger indicator LEDs do not illuminate,
replace the charger.
B. If the charger indicator LEDs illuminate, replace the
original Pak.