CONDENSATE

WATER INLET

d. Connect the connection tube included with the device to the water sup-
ply quick-release coupling on the device's rear panel, marked as WATER
IN. Submerge the other end of the tube in the water supply container.
The device is equipped with a water suction pump, there is no need to
position the water container above or on the same level as the device.
In order to secure the water supply tube, use the plug included in the
delivery and place the plug in the opening of the water supply container.
e. The wastewater formed after the water is turned into steam during the
HEPA
sterilisation process can be re-
moved using the tube provided,
which should be connected to
the port at the device's rear pan-
el, marked as WATER OUT. The
wastewater can be removed
directly to the sewerage or to a
special container intended for
wastewater. If using container,
place the tube end inside the
container and secure the inlet
with the plug provided. The
tube must not be submerged
in the wastewater.
f. The wastewater container or the sewerage drain must be located below
the device.
g. If using wastewater containers, we recommend using containers of the
same volume as those used for the deionised water. Emptying them
concurrently with replacing/filling the deionised water containers will
prevent overflow.
Correct positioning of cables in the water supply and wastewater
containers.
h. Leave 5 cm of space behind the device and 1 cm on each side from walls
or other elements in order to ensure sufficient ventilation.
i. The device should be positions in a way that ensures easy access to the
main switch located on the rear panel of the device.
j. Do not position the device near to washbasins or other places where it
could be poured with water - possible short-circuit.
k. Install the device in a well-ventilated room, away from heat sources and rooms
where mixtures of gases or liquids, and other hazardous agents may form.
l. Ensure the following environment conditions: operating temperature
range +5°C to +40°C/ relative humidity 0–90%, storage temperature
range from -20°C to +60°C/ relative humidity 0–90%.
Enbio S and Enbio PRO devices are designed for self-assembly by the
end user and do not require any special installation at the place of
use. The user is responsible for the correct installation of the device
on spot, according to this manual

3.1 Water quality

ENBIO sterilisers use demineralised or distilled water to form steam dur-
ing the sterilisation process. The total mineral content in the water used
for sterilisation must be lower than 10 ppm, or for conductance measure-
ments, lower than 15 μS/cm.
Standard tap water has hardness within the 2–3 mmol/l range and must
not exceed 5 mmol/l according to current regulations, making it unsuitable
for use in ENBIO sterilisers.
The table below presents the hardness and conductance parameters of
36