How to perform the test:
• Perform the test with an empty chamber, as per the EN 13060 standard.
• Place the test strip in the capsule. Read the instruction provided by the
test's manufacturer.
• Close the capsule.
• Place the test in the middle of the tray in the chamber.
• After completing the test, open the steriliser and remove the test.
WARNING! The package will be hot.
In order to interpret the test correctly, read the instruction provided by the
test pack manufacturer.
– Open the capsule and remove the test strip.
A part of the indicator strip did not change colour to the dark one due to
the presence of air in the capsule.
An insufficient change of colour of the indicator strip indicates the pres-
ence of air during the test cycle, caused by faulty operation of the steriliser.
POSITIVE RESULT
All fields of the indicator strip became
dark.
NEGATIVE RESULT
Part of the indicator strip did not change
color on the dark due to the presence of
air inside the capsule. Part of the indicator
strip did not change color on the dark due
to the presence of air inside the capsule.
5.4 Information menu
The information menu can be accessed by pressing the
be displayed.
You can also change the date and time.
field.
Here, information about the device
type, serial number, number of per-
formed processes, amount of free
memory available on the USB drive
for saving process data, and the
service menu
with the
process counters for sealing, filter
and the next service inspection can
In order to set the date or time, touch
the digits on the display. When a
field is selected, it starts to blink
and arrows used to
change values are displayed, up
and down
. This way you can
correctly set the date and time.
Pressing the number again confirms
it, and you can change another pa-
rameter.
We can choose the language in the
same way by clicking on the shortcut.
44