Replacement Parts; Periodic Inspections - ENBIO S Mode D'emploi

Masquer les pouces Voir aussi pour S:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 62
A message on the screen will notify the user when individual elements
should be replaced.
During regular operation, info screens concerning replacement of individ-
ual elements or required service inspection are displayed alternately. They
are described in detail in the „Warning messages and error codes" section.
In order to ensure efficient sterilisation and correct operation of the
device, it is recommended to observe the replacement dates for el-
ements subject to wearing down.
Pursuant to the medical products act, the manufacturer has speci-
fied the serviceability duration as 3 years or 6 000 cycles, provided
that regular servicing and timely service inspections are ensured.
After this period, the steriliser should be decommissioned.
This period has been estimated on the basis of pressure tests per-
formed on the pressure chamber (nonreplaceable components),
which withstood 6 000 cycles (corresponding to typical wear during
3 years) without plastic deformations.

6.1 Replacement parts

The following table includes elements subject to periodic replacement, and
elements subject to natural wear and tear. Replacement parts should be or-
dered directly from the manufacturer. Use of other replacement parts voids
the warranty and does not guarantee correct functioning of the device.
Name
Front seal
Bacteriological filter
Part no.
ST1-UL1
DZ0035
Connection/water supply tube
Connection/water supply tube
Rubber plug for the water
container
Rubber plug for the condensate
container

6.2 Periodic inspections

In order to ensure correct functioning of the ENBIO steriliser, it is recom-
mended to perform periodic service inspections and replace parts subject
to wearing down in accordance with the following schedule, and periodic
inspection of individual steriliser elements in accordance with the follow-
ing guidelines.
Name
Bacteriological filter
Connection/water removal tube
Plugs for water/condensate
containers
Element subject to inspection
Front seal
Bacteriological filter
Connection/water removal tube
Container plugs
ST1-HW1
ST1-HW2
ST1-KS1
ST1-KS2
Replacement frequency
every 1000 cycles or every 12
months
if damage is observed
if damage is observed
Inspection frequency
weekly or in the event of incor-
rect functioning – performed
by the user
weekly – performed by the user
weekly or in the event of incor-
rect functioning – performed
by the user
weekly – performed by the user
48