Replacing The Battery - Dentsply Sirona CEREC Primescan AC Notice D'utilisation

Masquer les pouces Voir aussi pour CEREC Primescan AC:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6 Maintenance

6.6 Replacing the battery

82
Replacing the battery
6.6
A
Service cover
B
Storage battery
C
2 fastening screws
1. Open the service cover on the back panel.
2. Remove the 2 fastening screws (C, cross-head screws).
3. Remove the storage battery.
4. Slide in the new storage battery using the guide rails and guide pins
up to the stop.
NOTE
Risk of fault during operation and defects to the system
If the storage battery is not screwed down, faults during operation and
defects to the system may occur.
➢ Always screw the storage battery down tightly.
5. Screw the storage battery in place with 2 fastening screws (C).
6. Put the service cover back in position and lock it.
Dentsply Sirona
D
2 guide rails
E
2 guide pins
67 15 481 D3696
D3696.201.01.03.02    04.2019