Dentsply Sirona
67 15 481 D3696
D3696.201.01.03.02 04.2019
Charging the battery
6.5
NOTE
Information on back-up cycles
The storage battery is designed for fully cable-free use of Scan,
Design and Grinding at a power requirement of 250 W for 60 minutes.
Approx. 2.5 hours are needed for full charging.
For back-up mode, which the user, for example, uses for an operation
lasting 10 minutes, in order to move the unit from door to door
(standby consumption of 100 W ), a charging time of 10 minutes is
needed. After around 1000 buffer cycles the capacity of the battery
fades due to the nature of the battery technology used.
The battery is permanently charged during operation on mains voltage.
For full charging, keeping the acquisition unit connected to the mains
voltage and the power switch on is sufficient. The PC does not have to
be switched on for the charging process.
NOTE
Reduced battery service life
If the battery is not charged over a long period of time, this
significantly reduces its service life.
➢ Always recharge the battery fully after buffer operation.
6 Maintenance
6.5 Charging the battery
81