6 Maintenance
6.1 Cleaning, disinfection and sterilization
68
9. Insert the metal holder with the sleeve.
10. Place the lid on the container and leave the sleeve at least 12
minutes in the disinfectant. While exceeding the recommended
immersion time may not cause damage to the sleeve, the sleeve
should be removed as soon as possible after the recommended
time.
Dentsply Sirona
67 15 481 D3696
D3696.201.01.03.02    04.2019