Dentsply Sirona CEREC Primescan AC Notice D'utilisation page 80

Masquer les pouces Voir aussi pour CEREC Primescan AC:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6 Maintenance
6.2 Calibrating the scanner
76
Measuring the position of the exit window
1. Mount the bottom side of the calibration set to the tip of the scanner.
2. Click on the "OK" button.
Ä The calibration process is continued.
Ä Once the calibration is complete, a message is displayed
indicating this.
3. Confirm the message by clicking the "OK" button.
Ä The CEREC Primescan / Primescan / CEREC Omnicam / Omnicam
scanner is calibrated.
Error message during calibration
The software indicates if an error occurs during calibration. If the
calibration process resulted in errors, restart the process.
End calibration
ü The software indicates that the calibration was completed
successfully.
➢ Click the "OK" button.
Ä The CEREC Primescan / Primescan / CEREC Omnicam /
Omnicam scanner is calibrated.
Dentsply Sirona
67 15 481 D3696
D3696.201.01.03.02    04.2019

Hide quick links:

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Primescan acCerec omnicam acCerec primescan aOmnicam ac

Table des Matières