5 Operation
5.3 Scanner guide
56
Procedure 2
1. Start occlusally on the distal tooth, tilt the scanner approx. 60° in an
oral direction and move it orally along the dental arch up to the
opposite distal tooth.
2. Guide the scanner occlusally from the distal tooth across the entire
dental arch back to the other side.
3. To complete the scan, tilt the scanner approx. 60° in a buccal
direction and move it buccally along the entire dental arch.
5.3.7
Concluding the optical impressions
ü The exposures are complete.
1. Click on the "Next" button.
Ä The virtual model is calculated and displayed in color.
Ä Beige brown sections highlight data material that is missing
from the calculated model.
2. If missing data emerges in the preparation area, carry out further
scans.
Change back to the "ACQUISITION" phase. Perform additional scans to
complete the model structure.
Dentsply Sirona
Operating Instructions
67 15 481 D3696
D3696.201.01.03.02 04.2019