Dentsply Sirona
Operating Instructions
67 15 481 D3696
D3696.201.01.03.02 04.2019
4.4.4
Switching the units on
NOTE
Do not put the unit into operation at low temperatures!
If you move the unit to the operating site from a cold environment,
condensation may form and result in a short circuit.
ü Install the unit at room temperature.
➢ Wait until the unit has reached room temperature and is
absolutely dry (for at least one hour)
Ä The unit is dry and can be put into operation.
CAUTION
Use only the supplied power cord
Use only the power cord supplied by Dentsply Sirona to connect the
acquisition unit to the power supply.
If the acquisition unit is switched on at the main switch, then it can be
switched on at the ON button. The monitor is switched on and off
automatically.
A
Main switch
1. Switch the acquisition unit on at the main switch.
2. Start the unit by holding down the On button (blue LED) on the right
at the back of the control console until a second vibration can be
felt. The color of the LED changes from blue to green.
NOTE
Possible data loss and PC malfunction:
Switching the acquisition unit off at the ON button during operation
may cause data loss and PC malfunctions.
➢ Always switch the unit off as described in the chapter "Switching
the units off [→ 42]".
4 Installation and startup
4.4 Initial startup
41