6 Maintenance
6.7 Replacing the O-ring
84
1. Slide the O-ring at one point into the O-ring groove.
2. Starting from this point, carefully use your thumb and index finger to
push the O-ring into the groove (in a similar way to pushing it out
but without any force).
3. Guide your thumb and index finger in such a way along the groove
until the O-ring is fully positioned in the groove.
NOTE
The O-ring can be damaged
Ensure that the O-ring fits in the groove without twisting in order to
prevent damage to the O-ring when fitting the sleeve.

IMPORTANT

Replacement O-rings
Replacement O-rings are available with the scanner or can be
ordered under REF 66 80 974.
Dentsply Sirona
67 15 481 D3696
D3696.201.01.03.02    04.2019