5.3.3
Lingual scan
ü The scanner is on the tooth that is positioned next to the
preparation in the distal direction.
1. Rotate the scanner to maximum 20° toward the lingual direction.
2. Guide the scanner over the entire lingual distance in the mesial
direction over the prepared tooth.
5.3.4
Approximal surface scan
Scan the approximal surfaces of the prepared tooth.
➢ Move the scanner in the occlusal direction to the prepared tooth.
Acquire the approximal surfaces in the distal and mesial direction.
5 Operation
5.3 Scanner guide
53