IEI
Work on electric, electronic and cryogenic components is limited to qualified personnel.
Master switch on shaker / mixer won't light
1. Check power cord
2. Check circuit breakers
Item won't shake (master switch lighted))
1. Timer run-down
2. Thermal motor circuit breaker triggered by motor overload
Remedy:
Wait about 20 minutes, decrease load applied on shaker plate.
3. Mechanical parts broken (humming motor noise) or electronic failure (no motor noise).
All shakers
space required
shaker plate
weight of item
ambient temperature
power dissipation
voltage / frequency
protective class
drive motor
TROUBLESHOOTING
SPECIFICATIONS
320 mm x 375 mm
258 mm x 290 mm
features anti-skid, detachable rubber plate with flanged
edges
approx. 8 kg
0°C to 50°C at 80% rel. humidity
approved for installation in gassing and conditioning cabinets
(make reference to temperature limits)
UNIMAX 1010 / PROMAX 1020
DUOMAX 1030 / POLYMAX 1040 = 25 W
230/240 V, 50/60Hz; 115V 50/60Hz optional
IP 40
condenser motor, electr. control, overheat protection
24
= 30 W