Interconnections
The following figure shows the TV 551/701 interconnections.
Figure 17
Earth (Ground) Connection
We recommend that you fit a separate earth (ground) conductor to the TV551/.
Use a separate insulated green/yellow conductor,AWG16 and use a M5 x 10
screw and shake proof washer (fitted to the earth hole on the pump) to secure
the earth conductor to the pump. The screw must be fixed with a 2 Nm torque.
The impedance between the pump-body and the earth connection point must
be < 0.1 Ω.
TV 551/701 Navigator User Manual / 87-900-934-01 (P)
TV 551/701 interconnections
Technical Information
Interconnections
275/328
17