Installation
Install this device indoors only, and in any case do not use the pump
in an environment exposed to atmospheric agents (rain, ice, snow),
dust, aggressive gases, or in explosive environments or those with a
high fire risk.
The Turbo-V controller must be positioned so that free air can flow through the
NOTE
holes.
WARNING!
If this instrument is not used as specified, the protection provided by the
equipment could be impaired. This instrument must be used in a normal
condition (in which all means for protection are intact) only.
During operation, the following environmental conditions must be
respected:
Do not remove the adhesive and protective cap before connecting the turbopump
CAUTION!
to the system.
TV 551/701 Navigator User Manual / 87-900-934-01 (P)
maximum pressure: 2 bar above atmospheric pressure
temperature: from +5 °C to +35 °C (see the diagram pressure-
temperature in the appendix "Technical Information")
relative humidity: 0 – 90 % (non-condensing)
Instructions for Use
Installation
251/328