Pump Used In Presence Of Magnetic Fields - Agilent Technologies TV 551/701 Navigator Notice De Mode D'emploi

Table des Matières

Publicité

Figure 58

Pump Used in Presence of Magnetic Fields

Magnetic fields induce eddy currents in the rotor of a turbomolecular pump
that tend to oppose to its rotation. The result is increased electrical power
consumption by the motor, most of which is dissipated in the rotor.
Since the rotor is not in contact with the stator the above power can leave the
rotor mainly by radia-tion and hence the rotor may be overheated while static
parts of the pump remain cool. This effect is strongly dependant from the
intensity, time function and distribution of the magnetic field. In general,
therefore, an increase in pump current can be expected.
If this increase is lower than 50 % of the current value drawn by the motor in
high vacuum opera-tion, no particular problem should be expected. However if
the effect is grater, than the case should be carefully reviewed by Agilent's
special-ist. As a matter of fact, in case of high magnetic fields, also important
forces might be generated and applied to the rotor.
TV 551/701 Navigator User Manual / 87-900-934-01 (P)
1
Purge gas line
2
Pressure regulator
3
Gas purge valve
4
Gas purge port
5
Forevacuum pump
6
Turbopump
Purge layout
Technical Information

Pump Used in Presence of Magnetic Fields

17
319/328