Test Function - Medela Vario 8 Mode D'emploi

Masquer les pouces Voir aussi pour Vario 8:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 22
Application
Vlll
ok?
OK
1. Check general condition of the device
• Is the overflow protection / bacteria filter present and OK?
(Test – see Chapter X)
All applications: overflow / weekly
Thoracic drainage: change of patient
Recommended for all other applications: Perform filter test
according to instructions for use after each patient
• Check seal of tubing connections
• Suction jars and lids – Check for cracks, brittle areas,
discolouration? – exchange, if necessary
• Mechanical overflow protection
must be open

2. Test function

a. Switch the device on
• AC/DC and c/i versions: before first use, charge the
rechargeable battery for five hours
• AC versions: plug in
• With AC/DC and c/i versions, select continuous pumping mode
b. Set maximum vacuum
Press knob to turn
c. Attach patient tubing and clamp tubing
d. Compare vacuum
+ 2000 m
not OK, see Chapter XI
3. Use
Select correct vacuum and suction mode according to
the particular application)
Press knob to turn
low
vacuum
– 7,0 kPa
– 7,9 kPa
+1000 m
– 8,5 kPa
+ 500 m
– 9,0 kPa
0 m
medium
high
vacuum
vacuum
– 43 kPa
– 59 kPa
– 48 kPa
– 66 kPa
– 70 kPa
– 51 kPa
– 55 kPa
– 75 kPa
EN