Cleaning
lX
Disconnect the device before cleaning
• Comply with in-house directives on hygiene
• Clean the device with a damp cloth and with non-aggressive cleaning material
• Suction jars, lids, suction bags and other accessories are to be handled in accordance
with the instruction-sheets provided
Test filter
X
Filter open!
1. Select maximum vacuum
2. Switch on the Vario
3. Read / compare the vacuum
If the mentioned values are exceeded, the filter must be replaced.
Xl
Troubleshooting
Insufficient
Check
vacuum
• Setting of the vacuum regulator
• Condition of the tubings
• Seal of the tubings connections
• That the overflow protection is functioning correctly
• If the filter is blocked
• Examine the suction jars, lids and seals for blockages / brittle areas
Motor not
Check
running
• Is mains/battery power present?
• Is the device switched on?
• Is the fuse ok?
c/i
Check the interval times
versions
• Check the jumper setting
• Correct position of the On/Off switch
If the fault cannot be corrected, contact the internal technical service
8
EN
Replace fuses according to the specification plate
Before changing the fuses, it is essential to
disconnect the pump from the mains supply!
Vario 8
Vario 18
low
medium
vacuum
vacuum
- 3 kPa
- 10 kPa
Vario 18
high
vacuum
- 10 kPa
On
Off
8 Secs.
4 Secs.
16 Secs.
8 Secs.
32 Secs.
16 Secs.