Standards
This electronic control conforms to
electrical safety standards in line with
standard IEC 60601-1 and those gov-
erning electromagnetic compatibility
in line with standard IEC 60601-1-2.
Electromagnetic compatibility
Caution: Portable RF communications
equipment (including peripherals such
as antenna cables and external anten-
nas) should be used no closer than 30
cm (12 inches) to any part of the de-
vice, including cables specified by the
manufacturer. Otherwise, degradation
of the performance of this equipment
could result. Declaration by the man-
ufacturer regarding electromagnetic
compatibility: refer to the tables on
pages 45-46-47.
Information
The technical specifications, illustra-
tions and dimensions contained in
these instructions are given only as a
guide. They may not be the subject of
any claim.
The manufacturer reserves the right to
make technical improvements to its
equipment, without amending these
instructions. For all additional infor-
mation, please contact Bien-Air Dental
SA at the address indicated on the
back cover.
Options
Converter 24/32-24
REF 1500580-001
The DMX2 Plus/DMX2 Pro is fed by 32
Vdc +/-10%. If your system only has a
24 Vac feed, we recommend that you
use this converter. This accessory en-
ables you to obtain optimum perfor-
mance from the MX2 system, by pro-
viding two stabilised voltages: 32 Vdc
(60W continuous, 130W peak) for the
MX2 motor feed.
Dual Motor Switch
REF 1500554-001
We recommend this module when
using 2 MX2 motors with a
Plus/DMX2 Pro board. It allows switch-
ing of the 3 motor phases, the 2 light
connections. The relays are switched
simultaneously and controlled by the
MUX Control input (24 Vdc). To con-
nect this electronic control to the MX2
system, please consult the wiring dia-
gram.
Disposal
This device must be recycled.
Electrical and electronic equip-
ment may contain dangerous sub-
stances which constitute health and
environmental hazards. The user must
return the device to its dealer or estab-
lish direct contact with an approved
body for treatment and recovery of this
type of equipment (European Directive
2012/19/EU).
Maintenance
Servicing
Never disassemble the device. For any
modification and repair, we recom-
mend that you contact your regular
supplier or Bien-Air Dental directly.
DMX2
Environment
Working
• Temperature:
+10°C (50°F) to +40°C (104°F)
• Relative humidity: 30% to 80%,
including condensation
• Atmospheric pressure:
700 hPa to 1060 hPa
Transport and storage
Environmental conditions for a period
of max. 15 weeks
• Temperature:
-25°C (-13°F) to +70°C (158°F)
• Relative humidity: 10% to 95%,
including condensation
• Atmospheric pressure:
500 hPa to 1060 hPa
Other precautions for use
The device must be used by a qualified
person in accordance with the current
legal provisions concerning industrial
safety, health and accident prevention
measures, and these working instruc-
tions. In accordance with these re-
quirements, the operators:
• must only use operating devices
that are in perfect working order; in
the event of irregular functioning,
excessive vibration, abnormal heat-
ing or other signs indicating mal-
function of the device, the work
must be stopped immediately; in
this case, contact a repair centre
that is approved by Bien-Air Dental;
• must ensure that the device is used
only for the purpose for which it is
intended, must protect themselves,
6