Guidance and manufacturer's declaration – electromagnetic immunity
The DMX2 Plus/DMX2 Pro is intended for use in the electromagnetic environment specified below. The customer or the user of the DMX2 Plus/DMX2 Pro should as-
sure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted distur-
3 VRMS
bances induced by
0,15 MHz – 80 MHz
RF fields
IEC 61000-4-6
6 VRMS in ISM bands
0,15 MHz – 80 MHz
80% AM at 1 kHz
Radiated RF EM
3 V/m
fields
80 MHz – 2,7 GHz
IEC 61000-4-3
80 % AM at 1 kHz
Proximity fields
Test freq. [MHz]
from RF wireless
385
communications
450
equipment
710, 745, 780
IEC 61000-4-3
810, 870, 930
1720, 1845, 1970
2450
5240, 5500, 5785
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and mobile field radios, amateur radios, AM and FM radio
broadcasts and TV broadcasts cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an elec-
tromagnetic site survey should be considered. If the measured field strength in the location in which the DMX2 Plus/DMX2 Pro is used exceeds the RF compliance
level mentioned above, the DMX2 Plus/DMX2 Pro should be observed to verify that it is operating normally. If abnormal operation is observed, additional measures
may be necessary, such as reorienting or relocating the DMX2 Plus/DMX2 Pro.
Compliance level
3 VRMS
0,15 MHz – 80 MHz
6 VRMS in ISM bands
0,15 MHz – 80 MHz
80% AM at 1 kHz
3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Immunity test level
Max. power [W]
1.8
27
28
2
0.2
9
28
2
2
28
2
28
0.2
9
Electromagnetic environment - guidance
Field strengths from fixed RF transmitters, as determined by
an electromagnetic
site survey a should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment marked
with the following symbol:
Distance: 0.3 m