The sterilisable products are supplied non sterile and need to be decontaminated and sterilised prior to first use and immediately after
each use. Use disinfectant products that comply with the instructions given by the manufacturer.
g
Anthogyr sterilisable medical products can withstand a maximum temperature of 138°C.
Remove the irrigation line and place it in a container for soiled medical devices (single use irrigation line).
E
Do not use the spray directly on the device to clean it; use disinfectant wipes instead.
10.2 Sterilisation cycle
1. Surgery
12. Autoclave/ Sterilisation
11. Packaging
10. Assembling
2. Purge
g
3''. Cleaning with disinfectant wipes
9. Drying
3. Sorting out of
the instruments
3'. Non-immersible instruments
Disinfectant device
8. Rinsing
IMPLANTEO
Dispose
of single
use accessories
4. Immersible instruments
5. Pre-disinfection
6. Rinsing + brushing
e
7. Ultrasonic bath
47
®