Emissions test
RF CISPR 11 emissions
Harmonic emissions
CEI61000-3-2
Voltage fluctuation /
Flicker twinkling
CEI61000-3-3
The IMPLANTEO® is intended to be used in the electromagnetic environment specified below.
The customer or user of the IMPLANTEO® must ensure that it is used in such an environment.
Immunity test
Electrostatic
discharges
CEI 61000-4-2
Rapid transition
into salvos
CEI 61000-4-4
Transitory excess
voltage
CEI 61000-4-5
Brief cut in voltage
and variations
in voltage in input
electricity supply
cables
CEI 61000-4-11
Magnetic field
for the frequency
of the electricity
network (50/60Hz)
CEI 61000-4-8
NOTE : U
is the alternative voltage before applying the test level
T
Compliance
Class B
Class A
Compliance
Directives and manufacturer's declaration – Electromagnetic immunity
Test level CEI 60601
± 6 kV on contact
± 8 kV in the air
± 2 kV for electricity
supply cables
± 1 kV for input/
output cables
± 1 kV between phases
± 2 kV between phases
and earth
<5 % UT
(>95 % cavity of UT)
for 0,5 cycle
40 % UT
(60 % cavity of UT)
for 5 cycles
70 % UT
(30 % cavity of UT)
for 25 cycles
<5 % UT
(>95 % cavity of UT)
for 5 s
3 A/m
Electromagnetic environment - directives
The IMPLANTEO® is suitable for use in all facilities,
including domestic facilities and those directly connected
to the low voltage public electricity network supplying
buildings for domestic use.
Compliance level
Electromagnetic environment - directives
Floors must be wooden, concrete or ceramic
± 6kV on contact
tiles. If floors are covered with synthetic
± 8kV in the air
materials, the relative humidity must be
at least 30 %.
± 2 kV for electricity
The quality of the electricity supply network
supply cables
must be that of a typical commercial or hospital
± 1 kV for input/output
environment.
cables
± 1 kV between phases
The quality of the electricity supply network
± 2 kV between phases
must be that of a typical commercial or hospital
and earth
environment.
<5 % UT
(>95 % cavity of UT)
for 0,5 cycle
The quality of the electricity supply network
40 % UT
must be that of a typical commercial or hospital
(60 % cavity of UT)
environment.
for 5 cycles
If the user of the IMPLANTEO® needs
70 % UT
continuous operation during cuts in supply,
(30 % cavity of UT)
it isrecommended supplying the IMPLANTEO®
for 25 cycles
from energy supply with no cut-out or with
a battery.
<5 % UT
(>95 % cavity of UT)
for 5 s
The magnetic fields at the frequency of the
electricity network must have the levels
3A/m
characteristic of a representative place located
in a typical commercial or hospital environment.