48
11. Abnormalities and errors
E
This section is intended to help users should a problem occur.
In the event of abnormal operation (screen that glistens, noisy pump motor, vibration in the micromotor), immediately
stop using the device and call Anthogyr After-Sales Dept. If it is not possible to locate or eliminate the problem using this
description, or if the disfunction symbols remain active, switch off the device and call Anthogyr After-Sales Dept. Under no
circumstances try to dismantle the control unit, motor or control pedal for the IMPLANTEO
Never try to repair it alone.
Abnormality screen
Dysfunction
The screen does not come on.
Analysis of disfunction
The pedal is not detected.
w Check that the pedal is properly connected to the box and check
the status of the connector.
w Press on the screen to make the safety signal go away and to go back
to the previous or initial screen.
The motor is not connected.
w Check that the motor is properly connected to the box and check
the status of the connector.
w Press on the screen to make the safety signal go away and to go back
to the previous or initial screen.
During the calibration process, the loading module is not connected to the contra-angle.
w Connect the loading module and validate to restart the operation.
w Press on the screen to make the safety signal go away and go back
to the previous screen.
Excess speed of the motor detected.
w Press on the screen to make the safety signal go away and go back
to the working screen.
Causes
w The device is switched off.
w The fuses are burnt.
w The wall supply socket is defective.
w The supply cable is defective.
.
®
Solutions
w Check that the device is properly connected
to the supply and that the On/ Off button
is in position 1.
w Replace the fuses.
w Move the device.
w Replace the cable.