40
6. Work screen
Select a user protocol and validate
C
When switching on the device, the welcome screen only stays active for 10 seconds before toggling to the work screen
(Fig.3)
associated with the active user / protocol.
When functioning, only the main data are displayed on the screen : Flow / Torque / Speed.
This is the description of the « work screen » according to the representation Fig.3.
Fig. 3
2
3
Press the centre of the screen
and maximum torque.
C
With the work screen, it is also possible to adjust the flow of peristaltic pump with a short press on button A of the pedal
(see section « 8 – Using the pedal » ).
6.1 Settings Flow / Speed / Torque
This is the description of the screen « settings Flow/ Speed / Torque » according to the representation
Fig. 4
1
6
8
10
6.2 Adding / Deleting / Replacing sequences
Press the icon
(Fig.1, rep.5)
to access the work screen (Fig.3).
1
Anthogyr
12
11
13
25
4
9
Rpm
10
35
5
N.cm
7
8
6
(Fig.3)
to have access to the settings of the 3 main data: irrigation flow, maximum speed
2
3
4
5
7
9
11
Press on buttons « + » and « - » or move the cursor on the bargraph.
(Fig.2, rep.1 or Fig.3, rep.9)
to access the sequence management screen (Fig.5).
1. Active User / Protocol and access to welcome screen.
2. Reduction ratio for active contra-angle and choice of it.
3. Activate or deactivate the brightness of the contra-angle.
4. Instantaneous speed of the tool if the motor is active or speed limit
when stopped.
5. Instantaneous torque if the motor is active or maxi. reached when
the motor is stopped.
6. Active sequence of all the sequences included in the active protocol.
7. Programmed torque limit.
8. Access to graphical display.
9. Edition of sequences of the current protocol.
10. Rotation direction of the motor and reversed direction.
11. Access settings.
12. Adjust the flow of the peristaltic pump.
13. Irrigation level.
Tactile zone 13, 4, 5, 12, 7 : Access settings Flow/ Speed/ Torque.
1. Flow of peristaltic pump.
2. Decrease the value.
3. Adjusted flow.
4. Increase the value.
5. Setting cursor.
6. Speed of micromotor.
7. Set maxi. speed.
8. Torque limitation of the micromotor.
9. Maximum reachable torque.
10. Cancel.
11. Validate.
Fig.4
: