The user must check the safety of operation and the condition of the device before each use.
When in use, national legal requirements and the following requirements in particular, must be complied with :
w work place safety regulations.
w accident prevention regulations.
To ensure that the Anthogyr product is operational at all times and to keep it in good working order, follow the maintenance,
disinfection and sterilisation instructions (see section « 10 – Cleaning, disinfection and sterilisation » ).
Only Anthogyr technicians or technicians trained by Anthogyr are authorised to repair Anthogyr products..
Check settings after maintenance.

2.3 Safety instructions

2.3.1 External interference
The device is certified compliant with current electromagnetic compatibility standards.
ATTENTION
E
Users should however ensure that no electromagnetic disturbance poses an additional risk.
So as not to disturb the proper operation of electromagnetic devices, it is recommended complying with the separation
distances for the device (section « 12 – Electromagnetic compatibility » ).
WARNING
E
The device is not designed to withstand shocks from an electrical defibrillator.

2.3.2 Electrical connection

DANGER (ELECTRIC SHOCK / ELECTROCUTION)
Non-compliant electrical connection
Check that the supply voltage is the same as that shown on the control unit.
E
Check the supply cable before each use.
Never stick or remove a plug from a wall socket with damp hands.
Do not connect USB peripheral having an external power supply.
Any other use or connection of devices to USB port may create new risks and may require additional risk analysis.
2.3.3 Environment
ATTENTION
Do not tilt the device.
E
Ensure that cables do not prevent free movement of individuals.
Do not use the device near ionising radiation.
The device must be connected to a nearby wall socket which must be easily accessible. The required local supply
cables include a socket supporting at least 125% of the nominal voltage for the equipment.
Use it on a maximum impedance network of 4 Ω.