Emc Table - Compex FIT 5.0 Mode D'emploi

Masquer les pouces Voir aussi pour FIT 5.0:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 43
8 . EM C TAB L E
The Compex Stimulator requires special precautions regarding EMC and needs to be installed and put into
service according to the information provided on the EMC in this manual.
All RF wireless transmission devices can affect the Compex Stimulator. The use of accessories, sensors,
and cables other than those specified by the manufacturer, may result in greater emissions or reduce the
immunity of the Compex Stimulator.
The Compex Stimulator should not be used beside or stacked on other equipment, if adjacent or stacked
use is necessary, one must check the correct functioning of the Compex Stimulator within the context of the
configuration used.
The Compex Stimulator is intended for use in the electromagnetic environment specified below.
The customer or user of the Compex Stimulator should ensure that it is used in this environment
EMISSIONS TEST
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ emission
oscillations IEC 61000-3-3
RECOMMENDATIONS AND DECLARATION BY THE
MANUFACTURER CONCERNING ELECTROMAGNETIC EMISSIONS
COMPLIANCE
Group 1
Class B
Class A
Not applicable
ELECTROMAGNETIC ENVIRONMENT
- GUIDE
The Compex Stimulator uses RF
energy only for its internal operation.
Consequently, its RF emissions are
unlikely to interfere with any adjacent
electrical device (radios, computers,
telephones etc.).
Compex Stimulator is suitable for use in
any establishment, other than a private
dwelling or a place connected directly
to the low voltage mains supply which
powers residential buildings.