EN
SYNCHRONIZATION PROBLEM
If the synchronization process was discontinued or cannot run successfully for any reason (remote control
disconnected, power failure, etc.) the remote control may, in some cases, display this screen.
• Reconnect the remote control to the computer and restart the synchronization process.
BEHAVIOUR OF THE MODULE'S LED
The LED alternately flashes green and red: the module is out of range or not recognized by the remote
control.
• Ensure the remote control is on.
• Ensure that the module and the remote control are less than 2 metres away from each other.
The LED is still red.
• Ensure the module is charged.
• Try to restart the remote control and modules.
• If despite this the LED is still red, contact customer service provided and approved by Compex.
The LED does not turn on.
• Ensure the module is charged.
• If despite this the LED still does not turn on, contact customer service provided and approved by Compex.
30