Télécharger Imprimer la page

Boston Scientific Precision Spectra SC-1132 Mode D'emploi page 12

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 32
Implantable Pulse Generator
7. Unscrew and remove the tunneling tool handle.
8. Grasp the tip of the tool with one hand while holding the straw in place with the other hand. Pull the
tunneling tool shaft out through the straw.
9. Push the lead or extension through the straw, then withdraw the straw.
10. Pull the proximal end(s) out of the exit point.
11. Wipe the proximal end(s) clean.
CAUTION: Do not tunnel splitter.
Note: If using the 2x8 Splitter and performing a permanent trial, the splitter tails may be tunneled to the
exit site.
Note: The following Codman Disposable Catheter Passers may be used in place of the Boston Scientific
tunneling tool:
REF 82-1515 (36 cm); REF 82-1516 (55 cm); REF 82-1517 (65 cm)
Note: When using a Codman Disposable Catheter Passer, tunnel from the midline incision to the IPG
pocket using the standard technique.
Implantable Pulse Generator
90970875-06 Rev A
8 of 285
Production
Boston Scientific Confidential. Unauthorized use is prohibited.
Spectra IPG Entrada 2.0 19L
92070451 A.1
Page 12 of 290

Publicité

loading

Ce manuel est également adapté pour:

Precision spectra system implantable pulse generatorSc-1132Precision spectra