Guidance and manufacturer's declaration - electromagnetic immunity
The Skin IQ MCM is intended for use in an electromagnetic environment specified below. The customer or user of the
Skin IQ MCM should assure that it is used in such an environment.
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1, At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2, These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Skin IQ MCM is used exceeds the applicable RF
compliance level above, the Skin IQ MCM should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the Skin IQ MCM.
b) Over the frequency range 150kHz, field strengths should be less than [V1] V/m.
12
IEC 60601 Test
Compliance Level
Level
3Vrms
150K - 80 MHz
150K - 80 MHz
3 Vrms
80 MHz - 2.5 GHz
80 MHz - 2.5 GHz
Electromagnetic Environment Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the
Skin IQ MCM, including cables, than the
recommended separation distance calculated from
the equation application to the frequency of the
transmitter.
3 Vrms
Recommended Separation Distance
Battery Operated Device
3 Vrms
3.5
d=[ — ]
E
1
7
d=[ — ]
E
1
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey (see
note a) should be less than the compliance level in
each frequency range (see note b).
Interference may occur in the vicinity of equipment
marked with the following symbol:
P 80 MHz to 800 MHz
P 800 MHz to 2.5 GHz