Installation
See the Quick Reference on pages 2 and 3 for illustrations of the following procedures.
Failure to properly secure the coverlet to the existing surface may lead to
patient or user injury or equipment damage.
1.
Remove coverlet from shipping bag.
2.
Remove all covers and sheets from the existing pressure redistribution surface.
3.
Place the coverlet on top of the existing surface, ensuring the foot graphic on the
coverlet is at the foot end of the bed.
4.
Connect the power supply to the fan located underneath the foot end of the
coverlet.
5.
Ensure the cord that runs from the fan to the power supply is placed on the floor
under the bed. Improper placement of the cord could cause injury.
6.
Pull the coverlet over the pressure redistribution surface by stretching it over each
corner securely. Do not trap the power cord between the coverlet and mattress.
7.
Smooth any wrinkles on the coverlet.
8.
Secure the coverlet to the bed by using the hook-and-loop straps, located on the
underside of the coverlet.
9.
Ensure strap placement does not interfere with the operation of the bed functions.
Failure to do so could result in patient injury or equipment damage.
10.
Write the therapy start date on the law tag at the foot end of the coverlet to track
use. Failure to track duration of use may void warranty.
11.
Confirm there are no sharp objects in the immediate area which may damage the
coverlet.
12.
Connect the power supply to the power cord.
13.
Connect the power cord to a properly grounded electrical outlet and confirm
outlet has power. Verify that the electrical outlet can be easily accessed when
disconnecting the device from mains power. Ensure power supply and cord are
properly stored on the floor beneath bed.
14.
Move hand to the end of the bed and press the foot-end of coverlet to feel fan
vibration. A low humming sound indicates the fan is working.
EN
7