WEEE - Waste Management
The device and its accessories must be dis-
posed of separately as electronic waste in ac-
cordance with your local authorities.

Weight and Dimensions

 Weight: 10.9kg
 Dimension: 81x250x275mm (HxWxD)

Safety Checks

The following safety checks should be
conducted by qualified personnel at least once
a year and in the event of repair:
 Inspection for visible damage to device.
 Inspection of mains cord and connecting
cables.
 Possible: Measurement of insulation
resistance.
 Measurement of leakage currents.
 Measurement of resistance of protective
earth conductor.

Classification

Type of protection against electric shock:
 Class I: Equipment in which protection
against electric shock does not rely on basic
insulation only, but which includes an
additional safety precaution in that means
are provided for the connection of the
equipment to the protective earth conductor
in the fixed wiring of the installation in
such a way that accessible metal parts
cannot become live in the event of a failure
of the basic insulation.
Method(s) of sterilization:
 Device not intended to be sterilized
Degree of patient vicinity:
 Device suitable for use in patient environ-
ment.
Degree of protection against electric shock:
 Enclosure protectively earthed (treated as
applied part Type B).
Degree of protection against harmful ingress
of water:
 IP20: Ordinary equipment (enclosed
equipment without protection against
ingress of water).
Degree of safety of application in the presence
of a flammable anesthetic mixture with air or
with oxygen or nitrous oxide:
 Equipment not suitable for use in the
presence of such a mixture.
Mode of operation:
 Continuous operation at normal load and
Momentary operation at high load.
 Momentary operation at high load is used
for short time printing with high power
laser printers.
 Momentary operation at high load is
possible for up to 60 sec per hour at
elevated temperatures.
 Operation at high load over prolonged time
may cause thermal cut out. After cooling
down (hours) internal thermal fuse will
reset and normal operation can be resumed.

Accessories

 4 finger screws w. star washers (for cart
fixation)
 4 plastic/rubber feet (for desktop setup)
 2 fuses (as spare parts)