13. EMC / Technical description
The Swing maxi breast pump needs special precautions regarding EMC (Electro-
magnetic compatibility) and needs to be installed and put into service according to
the EMC information provided in these instructions for use.
Wireless communications equipment such as wireless home network devices,
mobile phones, cordless telephones and their base stations, walkie-talkies can affect
the electric breast pump Swing maxi and should be kept at least a distance 1.0 m
away from the equipment.
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 1)
Electromagnetic emissions
The electric breast pump Swing maxi is intended for use in the electromagnetic environment
specified below. The customer or the user of the electric breast pump Swing maxi should
assure that it is used in such an environment.
Emission tests
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
116
IEC 61000-3-3
Warning – The electric breast pump Swing maxi should not be used adjacent
to or stacked with other equipment. If adjacent or stacked use is necessary,
the electric breast pump Swing maxi should be observed to verify normal operation in
the configuration in which it will be used.
Electromagnetic environment –
Compliance
guidance
The electric breast pump Swing maxi uses RF
energy only for its internal function. Therefore,
Group 1
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
Class B
The electric breast pump Swing maxi is suitable
for use in all establishments, including domestic
Class A
establishments and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic pur-
poses.
Complies