Calibration and Autodiagnosis
46
In some cases, a selection menu
may appear for geometries. Select
the desired geometry and press on
the scroll wheel.
A warning message appears. You
can cancel out of this process by
pressing the operate button.
If you press the scroll wheel, you
will continue with the process of
changing calibration values.
In the next display you can enter
new calibration values.
After you have entered the new
value, you can again abort the
process with operate.
If you confirm the new value by
pressing the scroll wheel, the value
will be accepted.
After you have changed all
necessary values, you should
recalibrate the measurement
device as usual.