7
Initial operation
Remove the transportation safety device from the bottom of the housing, see sheet "Transportation safety
device"
Position the centrifuge in a stable and level manner in a suitable place. During set-up, the required safety
margin of 300 mm around the centrifuge is to be kept according to EN / IEC 61010-2-020.
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous
substances or objects may be within the safety margin of 300 mm around the centrifuge.
Substances can be discharged through the ventilation opening on the rear side of the
centrifuge. The device is to be placed in such a way that the air stream is not directed towards
people.
Do not place any object in front of the ventiduct.
Keep a ventilation area of 300 mm around the ventiduct.
Check whether the mains voltage tallies with the statement on the type plate.
Connect the centrifuge with the connection cable to a standard mains socket. For connection ratings refer to
Chapter "Technical specifications".
Turn on the mains switch. Switch position "".
The last used centrifuge data will be displayed.
Open the lid.
Remove the transportation safety device in the centrifuge, see sheet "Transportation safety device".
8
Opening and closing the lid
8.1
Opening the lid
The lid can only be opened when the centrifuge is switched on and the rotor is at rest. If it cannot be
opened under these circumstances, see the section on "Emergency release".
Rotate the turning handle on the front panel to the left. The symbol " " (lid open) illuminates in the rotation
indicator
.
Open the lid.
8.2
Closing the lid
Do not bang the lid shut.
Place the lid and lightly press down the front edge of the lid.
Turn the handle on the front panel to the right . The symbol " " (lid closed) illuminates in the rotation indicator
30/70
.